Lesson 1: Introduction to ISO 10993
Use this assistant to build a Biological Evaluation Plan (BEP) aligned with ISO 10993-1. We start with your device description, then map contact type/duration and materials to a risk-based test plan. Prefer chemistry-first (ISO 10993-18/-17) where appropriate.
How to use (3 minutes):
Enter Device basics and choose high-level category.
Select Body contact and duration.
Tick relevant materials and processes (e.g., EO sterilization).
Review risk flags and the automatically generated test plan with rationales.
Export/print your BEP summary.
Lesson 1: Scope & Contact Mapping
Select the tissue/organ contact and duration. This drives the baseline test matrix.
Why this matters (risk-based approach)
ISO 10993 emphasizes selecting tests based on nature of contact and duration. Long-term, deeper, or blood contact typically increases need for systemic toxicity, genotoxicity, implantation, and chemical characterization with toxicological risk assessment.
Lesson 2: Materials, Processes & Chemistry-First
Tick applicable materials and processes. The assistant will propose chemistry-first (ISO 10993-18/-17) and targeted biological testing.
Reduce unnecessary animal testing by using extractables/leachables and toxicological risk assessment when scientifically justified.
If yes, plan for medicinal-substance characterization and benefit-risk alignment with device biocompatibility.
Lesson 3: Risk Flags & Chemical Characterization
The assistant scans for risk flags that shape your plan (e.g., long-term implant, blood contact, EO residuals, degradation products).
Chemical characterization (10993-18) & toxicological risk assessment (10993-17)
Identify and quantify extractables/leachables from relevant test articles (per 10993-12 extraction) under clinically relevant conditions. Compare patient exposure to toxicological thresholds (TTC/PDE) and justify acceptance or additional testing. Degradation product standards: polymers (-13), ceramics (-14), metals (-15), and toxicokinetics (-16) where applicable.
Suggested ISO 10993 Test Plan
Generated from your inputs. Each item includes the ISO part and a brief rationale. Use the toggles to tailor, then export.
Clinical evaluation & PMCF link-up
For novel materials, borderline chemistry, or long-term implants, align with a Clinical Evaluation and (if applicable) Post-Market Clinical Follow-up to confirm safety signals remain acceptable in real-world use.
Plan Summary & Export
Risk tier
—
Based on depth/duration
Chem-first fit
—
Suitability estimate
Planned tests
0
# selected
Disclaimer
This assistant provides educational, risk-based guidance. Final test selection must follow ISO 10993-1 biological evaluation principles, professional judgment, and applicable regulatory expectations for your device and market.